DOI: 10.3290/j.qi.a37134, PubMed ID (PMID): 27834420Pages 33-39, Language: EnglishHorwitz, Jacob / Machtei, Eli E. / Zigdon-Giladi, HadarObjective: The purpose of the study was to assess the clinical accuracy of a novel open-lattice-frame implant positioning system by evaluating the deviation between planned and actual implant positions on pre- and postoperative cone beam computed tomography (CBCT) images of patients treated with dental implants. Secondary aims were to record surgical time from first incision to end of implant surgery and start of suturing, and to record patient and surgeon satisfaction from the implant surgical procedure. Pre- and postoperative CBCT of 10 patients receiving 18 dental implants were used to generate implant angular deviations and implant shoulder and apex deviations. Mean angular deviation was 2.96 ± 1.31 degrees (range 0.75 to 5.60 degrees). Mean shoulder and apex deviations were 1.07 ± 0.49 mm (range 0.38 to 1.85 mm) and 1.35 ± 0.57 mm (range 0.52 to 2.19 mm), respectively. Mean time from first incision to start of suturing was 24 ± 7 minutes. Patient satisfaction ranged between 8 and 10, and surgeon satisfaction regarding the procedure ranged between 9 and 10, on a scale of 1 to 10. The open-lattice-frame implant positioning system provided adequate accuracy, and may aid in improving patient-related and surgeon-related outcomes by improving surgical accuracy and shortening surgical treatment duration.
Keywords: accuracy, guided surgery, implant, planning