Purpose: To evaluate the effect of different abutment screw retightening procedures on the screw stability between the abutment and implant.
Materials and Methods: External hexagon connection (EHC) and internal trichannel connection (ITC) implants were employed in this study. Each abutment screw was used for the following specific applications (n = 5 for each group): group 0 = abutment was tightened and retightened with a 10-minute interval; group 1 = abutment was tightened and retightened with a 10-minute interval, then loaded for 500,000 cycles; group 2 = abutment was loaded for 83,000 cycles during tightening and retightening, then loaded for 500,000 cycles; group 3 = abutment was loaded for 167,000 cycles during tightening and retightening, then loaded for 500,000 cycles; and group 4 = abutment was loaded for 250,000 cycles during tightening and retightening, then loaded for 500,000 cycles. Both tightening and retightening torques were set at 35 Ncm, and dynamic oblique loading between 0 N and 200 N was applied on the abutment in all groups. The settling values of the abutments after retightening and the abutment removal torque values (RTVs) after testing were measured.
Results: No significant differences in settling values were found between EHC and ITC. There were significant differences in the RTVs of EHC implants, but ITC implants showed no difference in RTV with different prosthetic retightening applications.
Conclusion: The retightening application affected the joint stability of EHC implants but did not affect the settlement and joint stability of ITC implants.