Pages 419-433, Language: EnglishBinonThe design and engineering characteristics of the Spline dental implant were investigated. Compressive strength were evaluated. The results of the mechanical test indicated that this implant/abutment complex is mechanically stable, has minimal rotational movement, improved resistance to screw loosening, and good interface fidelity. The data for the new design are compared with published data for other implant/abutment connections.