Pages 21-24, Language: EnglishCarlsson, Lars / Röstlund, Tord / Albrektsson, Björn / Albrektsson, TomasThe ability to resist removal torques was compared for two different surface textures of screw-shaped commercially pure titanium implants. It was found that the removal torque for rough-surfaced implants was significantly higher in comparison to smooth-surfaced implants after 6 weeks of implantation time in rabbits.
Keywords: biomechanical test, electropolished implants, osseointegration, surface texture, titanium implants