DOI: 10.11607/prd.2082, PubMed ID (PMID): 26901305Pages 274-280, Language: EnglishCalcaterra, Roberta / Di Girolamo, Michele / Mirisola, Concetta / Baggi, LuigiScrew loosening can damage the interfaces of implant components, resulting in susceptibility to contamination of the internal parts by microorganisms. The aim of this study was to investigate the impact of abutment screw retightening on the leakage of two different types of bacteria, Streptococcus sanguinis and Fusobacterium nucleatum, and of the yeast Candida albicans. Two types of implantabutment systems with tube-in-tube interfaces were tested. Groups A and B each used a different type of system that consisted of 20 different pieces that were assembled according to the manufacturer's torque recommendations; four samples in each group were closed just one time, four samples three times, four samples five times, four samples seven times, and four samples nine times. The implants of groups A and B were contaminated with 0.1 μL of microbial solution just before being assembled for the last time to minimize the possibility of contamination. Results showed a direct correlation between the number of colony-forming units grown in the plates and the closing/opening cycles of the implant-abutment systems. Within the limitations of this study, the results indicate the possibility that repeated closing/opening cycles of the implant-abutment unit may influence bacterial/yeast leakage, most likely as a consequence of decreased precision of the coupling between the abutment and the internal part of the dental implant. These findings suggest that a one-time abutment technique may avoid microbiologic leakage in cases of implant-abutment systems with tube-in-tube interfaces.