Pages 522-528, Language: EnglishBasten, Christoph H.-J. / Nicholls, Jack I. / Daly, Colin H. / Taggart, RayThe load fatigue performance of the CeraOne and EsthetiCone implant systems was investigated. Three torques (standard torque, standard torque + 20%, and standard torque - 20%) were used to tighten the screws. A rotating-beam fatigue model was used. A 70-N load was applied to the specimens at a 30-degree angle to the long axis. Two-way analysis of variance revealed (1) no statistically significant difference between the three torques for each system, and (2) a statistically significant difference between the implant systems. Implant failure occurred in 10 of 15 CeraOne specimens. In the EsthetiCone system, only abutment screw failure occurred.
Keywords: implant, load angle, load fatigue, torque