Seiten: 217-222, Sprache: EnglischLopes, Hélio P. / Vieira, Márcia V. B. / Elias, Carlos N. / Siqueira jr., José F. / Mangelli, Marcelo / Lopes, Weber S. P. / Vieira, Victor T. L. / Rocha, Edilene AndradeObjectives: This study evaluated the influence of reciprocating movement on the fatigue life of root canal instruments subjected to static and dynamic tests.
Materials and methods: The rotary NiTi instruments used in this study were WaveOne and ProTaper F2 (both from Dentsply Maillefer, Ballaigues, Switzerland). The instruments were initially subjected to a cantilever-bending test, then to static and dynamic fatigue tests. WaveOne instruments were operated in reciprocating movement and ProTaper in continuous rotation.
Results: The mean bending resistance values (maximum load in grams) of the instruments were as follows: WaveOne = 544.8; ProTaper F2 = 555.4. No significant difference was observed between the instruments tested (P > 0.05). The mean time to fracture (in seconds) of the instruments subjected to static and dynamic tests were as follows: WaveOne = 93.5 (static) and 267.6 (dynamic); ProTaper F2 = 57.8 (static) and 94.1 (dynamic). Student's t test revealed significant differences in all tests (P 0.05).
Conclusion: The findings revealed that instruments operated in a reciprocating movement and in the dynamic testing model presented prolonged fatigue life.
Schlagwörter: dynamic test, fatigue fracture, fatigue resistance, flexibility, reciprocating movement